The function of the portable sampling system is automatic, but sample pump and valve can be controlled also manually. In the case of a power failure or if the temperature (pump, lines, sample cell) is below setting, the automatic 3-way valve switches sample gas to zero gas to prevent condensation. Sample pump cannot be switched on before all temperatures have reached the setting. In addition, the zero calibration of the FTIR gas analyzer can be done automatically with the portable sampling system. As an option, the sampling system can be equipped with a sample probe and/or heated lines. The maximum length for the heated line is 19 m + 1 m with 230 VAC and 9 m + 1 m with 115 VAC power supply. There is also an optional integrated O2 sensor that supplements the capabilities of the FTIR gas analyzers.
Description
Portable Sampling System The portable sampling system has been designed .for portable emission monitoring measurements The portable sampling system is used for on-site measurements. It can be used for measuring trace concentrations of pollutants in wet, corrosive gas streams. The sample gas can be measured undiluted and without drying since the sample pump, heated filter and valve are located in a module that is heated to 180 ºC. From the sampling system the gases can be directed into FTIR gas analyzer.
The portable sampling system includes power connections and temperature controllers for heated lines and heated module. The portable sampling system is connected to an external PC through FTIR gas analyzer and can be controlled .by software